Product Details
Disinfecting Type:
No
Properties:
CO2 Monitor
Size:
2.40 M
Stock:
Yes
Shelf Life:
1years
Material:
TPU
Quality Certification:
CE/ISO
Instrument Classification:
Class II
Safety Standard:
ISO13485
Payment & Shipping Terms
Minimum Order Quantity
1pcs
Packaging Details
PE bag, foam and box
Delivery Time
About one week
Payment Terms
T/T
Product Description

Sidestream ETCO2 For main brands

 

ETCO2 Sensor Catalogue.pdf

 

CO2 Sampling Lines Catalogue(1).pdf

 

Project Specifications
Principle of Operation Non-dispersive infrared(NDIR) single beam optics, dual wavelength, no moving parts.
Initialization Time Capnogram displayed in less than 3 seconds,full specifications within 3 minutes(At 25°C).
CO2 Measurement Range 0 - 114 mmHg; 0 - 15 %; 0 - 15.2 kPa.
CO2 Rise Time Rise Time Less than 80 ms - Adult Reusable or Single-Patient-Use Airway Adapter Less than 80 ms - Infant Reusable
or Single-Patient-Use Airway Adapter.
CO2 Resolution 0. 1mmHg ; 0 to 38 mmHg.; 0.25 mmHg ; 39 to 114mmHg.
CO2 Accuracy 0 - 40mmHg ±2 mmHg.; 41 - 76mmHg ±5% of reading.; 77 - 114 mmHg ±8% of reading.
Above 80 breaths per minute ±12% of reading. * Note: Gas emperature at 25°C.
CO2 Stability Short term drift: Drift over four hours shall not exceed 1 mmHg maximum.
Long term drift: Accuracy specification will be maintained over a 120-hour period.
Sampling Frequency 100Hz.
Respiration Rate Range 3 to 150 breaths per minute.
Respiration Rate Accuracy ±1 breath.
Calibration No routine user calibration required. An airway adapter zero is required when changing to a different style of airway
adapter.
Airway Adapters Single-patient-use or reusable, < 5 cc dead space (adult), < 1 cc dead space infant.
Temperature Operating: 0 to 40°C; Storage: -40 to 70°C; Transport: -40 to 70°C.
Humidity Operating: 10 to 90% RH, non-condensing; Storage: 10 to 90% RH, non-condensing.;
Transport: 10 to 90% RH, non-condensing.
Atmospheric Pressure Operating: 400-800 mmHg; Storage: 400-800 mmHg; Transport: 400-800 mmHg.
Water Resistance IPX4 - Splash-proof - Sensor only.
Voltage Requirements 5VDC ± 5 %.
Power Rating <1.0W Typical; <1.8 W Warm-up.
Sensor Weight <85g.
Sensor Size Deep*Wide*High:48mm * 24.3mm * 35.5mm.
Cable Length 2.4m.

 

 